Webinars and Sponsored Roundtables — Register Now
Thursday, August 20, 2026. 12 PM-1:30 PM ET
Hear from an expert in HER2 IHC whose work has contributed to current scoring guidelines.
Webinar presenter Josef Ruschoff, MD, PhD, Former Chief Medical Officer and Co-Founder of Targos, Senior Consultant of Discovery Life Sciences, Kassel, Germany.
Moderated by: Bob McGonnagle, Publisher, CAP TODAY
To be rescheduled to a future date
Learn about digital pathology technology that is future-ready, yet practical for today’s
laboratory needs.
Webinar presenters Scott Hammond, Senior Systems Consultant, Digital Pathology Division, Wexner Medical Center, Department of Pathology, and Ursula Hofer, Imaging Technologist, Pathology Digital Imaging Lab, Wexner Medical Center, Department of Pathology, and Sandra Banky, PA(ASCP), Director of Operations, Wexner Medical Center, Department of Pathology.
Moderated by: Bob McGonnagle, Publisher, CAP TODAY
Interactive Product Guides
Product Spotlight
Abbott
Alinity h-series
| Company | Abbott |
|---|---|
| Name of analyzer | Alinity h-series |
| Contact | James Pickering |
| City, State | Lake Forest, IL |
| Phone | 800-323-9100 |
| Website | |
| Analyzer application | diagnostic and research use |
| Dimensions of analyzer (H × W × | 59 × 24 × 36 in./— |
| Country where analyzer was | U.S./Singapore |
| First year analyzer was sold in U.S./Approximate No. of units installed in U.S.* | 2023/25 |
| First year analyzer sold outside | 2017/975 |
| Foreign locations where company | Brazil, Canada, Italy, Japan, Saudi Arabia, Spain, U.K., Vietnam, more (installed in the aforementioned countries) |
| Human languages (other than | Brazilian Portuguese, French, German, Italian, more |
| Analyzer list price | — |
| • Analyzer acquisition options | purchase, trade in, lease, reagent rental agreement |
| Training included with purchase/Total time for basic training per operator | 15 min. daily; 30 min. weekly |
| • Follow-up training available | yes (extra charge) |
| Parameters that analyzer tests and reports | %&# eosinophils, %&# immature granulocytes, %&# lymphocytes, %&# monocytes, %&# neutrophils, %&# basophils, %&# reticulocyte count, Hb, Hct, MCH, more |
| Types of testing FDA cleared or approved for analyzer | — |
| Types of body fluids FDA cleared or approved for analysis on analyzer | — |
| Analytical measurement range for body fluid | — |
| Precision for body fluid | — |
| Accuracy for body fluid | — |
| Automated differential performed for body fluids | — |
| Tests available for research use only | WBC viability fraction, % microcytic RBCs, % macrocytic RBCs, % hypochromic RBCs, % hyperchromic RBCs, more |
| Differential methods used | advanced MAPSS technology |
| Analytical measurement range: • WBC count/RBC count | WBC: 0.04–447.00 × 103 cells/µL/RBC: 0.01–8.08 × 106 cells/µL |
| Precision: • WBC count/RBC count | WBC: ≤3.5% CV @ >4.0 × 103 cells/µL/RBC: ≤1.5% CV @ >4.00 × 106 cells/µL |
| Accuracy: • WBC count/RBC count | — |
| Interfering substances: • WBC/RBC | cryoglobulin, cryofibrinogen, fragile WBCs, nonviable WBCs, neutrophil aggregates, hemoglobin C crystals, NRBCs, platelet clumps or aggregates/RBC autoagglutinins, cold agglutinins, giant platelets, RBC fragments |
| • Hemoglobin/Platelets | hemoglobin C crystals/RBC autoagglutinins, cold agglutinins, cryoglobulin, cryofibrinogen, giant platelets, platelet clumps or aggregates, platelet satellitism, RBC fragments, more |
| • Hematocrit/Reticulocytes | RBC autoagglutinins, cold agglutinins, giant platelets, platelet clumps or aggregates, more/RBC autoagglutinins, cold agglutinins, babesiosis, malaria, Heinz bodies, more |
| • Automated differential | cryoglobulin, cryofibrinogen, fragile WBCs, nonviable WBCs, neutrophil aggregates, hemoglobin C crystals, more |
| Throughput: max. CBCs per hour/Max. CBCs with differentials per hour | 119/119 |
| • Max. No. of slides processed per hour if stainer included | 65 |
| FDA-cleared or -approved sample container types (besides potassium EDTA) | none |
| Min. specimen volume open mode/Closed mode/Microsample capability | ≤100 µL/≤100 µL/no |
| Types of patient data archived/Max. No. of archived patient data accessible when analyzer online | patient demographics, previous test results, graphics, more (customizable)/— |
| Analyzer performs delta checks | yes |
| Parameters for which flags may appear | morphological flags, including left shift, blasts, variant lymphocytes, platelet clumps, RBC fragments, more |
| Analyzer recognizes and holds test results | yes, for follow-up, confirmatory testing, and potential test rerun (parameters for flags to hold samples defined by client or vendor) |
| Scattergram display uses cell-specific color | yes |
| Histogram display uses color with thresholds | yes |
| LIS interface formats supported | ASTM 1394-91, ASTM 1381, HL7 version 2, HL7 version 3 |
| • Middleware interface | yes (can use company’s middleware or third-party middleware or connect directly to LIS) |
| • Information transferred to LIS via middleware or LIS interface | numeric results, flagged results, histograms, scatterplots, patient demographics, patient orders, more |
| Analyzer transmits LOINC codes for results to LIS or middleware | no |
| Lab automation system or workcells to which analyzer has a direct interface | Abbott Diagnostics |
| Barcode symbologies read on specimen tube | Codabar, Code 39, Code 128, Interleaved 2 of 5 |
| • Analyzer allows barcode placement per CLSI standard Auto02-A2 | yes |
| No. of onboard reagents required for standard specimen analysis | 3 |
| • Reagents ready to use/Reagent tracking method | yes/barcode reader |
| Onboard diagnostics for troubleshooting | yes |
| • Manufacturer can perform diagnostics via remote access | yes |
| Warranty provided/Cost of annual service contract (24 h/7 d) | one-year warranty/variable |
| Notes | — |