Webinars and Sponsored Roundtables — Register Now
Tuesday, July 21, 2026, 11:00-11:30 AM CT
- Explain how transparency and manufacturer partnerships improve quality, consistency, and decision-making confidence in specimen management.
- Evaluate blood collection tubes beyond cost and commodity assumptions, incorporating clinical impact and risk into decision-making.
- Assess the potential risk points when using a blood collection device that has not been cleared for a specific purpose.
Roundtable presenters Nick Fingland, PhD, PMP, Senior Director, R&D Operations and Science, BD, and Chris Farnsworth, PhD, D(ABCC), Section Head of Clinical Chemistry, Professor of Pathology and Immunology, Washington University School of Medicine.
Moderated by: Bob McGonnagle, Publisher, CAP TODAY
Wednesday, July 29, 2026, 1:00-2:00 PM ET
Learn about digital pathology technology that is future-ready, yet practical for today’s
laboratory needs.
Webinar presenters Scott Hammond, Senior Systems Consultant, Digital Pathology Division, Wexner Medical Center, Department of Pathology, and Ursula Hofer, Imaging Technologist, Pathology Digital Imaging Lab, Wexner Medical Center, Department of Pathology, and Sandra Banky, PA(ASCP), Director of Operations, Wexner Medical Center, Department of Pathology.
Moderated by: Bob McGonnagle, Publisher, CAP TODAY
Interactive Product Guides
Product Spotlight
Sysmex Corp.
XN-350, XN-450, XN-550
| Company | Sysmex Corp. |
|---|---|
| Name of analyzer | XN-350, XN-450, XN-550 |
| Contact | Madelaine Dintelman |
| City, State | Lincolnshire, IL |
| Phone | 888-879-7639 |
| Website | |
| Analyzer application | diagnostic use |
| Dimensions of analyzer (H × W × | XN-550: 17.7 × 17.7 × 26 in./varies by configuration |
| Country where analyzer was | Japan/Japan |
| First year analyzer was sold in U.S./Approximate No. of units installed in U.S.* | 2017/>2, 500 |
| First year analyzer sold outside | 2015/— |
| Foreign locations where company | — |
| Human languages (other than | — |
| Analyzer list price | — (slide maker stainer sold separately) |
| • Analyzer acquisition options | purchase, trade in, lease, cost per reportable contract |
| Training included with purchase/Total time for basic training per operator | 2 min. daily; 15 min. weekly |
| • Follow-up training available | yes (no extra charge) |
| Parameters that analyzer tests and reports | %&# eosinophils, %&# immature granulocytes, %&# lymphocytes, %&# monocytes, %&# neutrophils, %&# basophils, %&# reticulocyte count, Hb, Hct, MCH, more |
| Types of testing FDA cleared or approved for analyzer | blood count analysis, body fluid analysis |
| Types of body fluids FDA cleared or approved for analysis on analyzer | CSF, peritoneal/ascites, pleural fluid, synovial fluid (on XN-L 50 series only) |
| Analytical measurement range for body fluid | WBC count: 0.004–10.000 × 103/µL; RBC count: 0.002–5.000 × 103/µL; TNC count: 0.004–10.000 × 103/µL |
| Precision for body fluid | WBC count: ≤30.0% (0.005–0.015 × 103/µL), ≤20.0% (0.016–0.030 × 103/µL), ≤15.0% (0.031–0.050 × 103/µL); RBC count: ≤40.0% or ≤ 0.007 × 106/μL (0.003–0.050 × 106/μL) max.–min.; TNC count: ≤30.0% (0.005–0.015 × 103/μL), ≤20.0% (0.016–0.030 × 103/μL), ≤15.0% (0.031–0.050 × 103/μL) |
| Accuracy for body fluid | WBC count: r = 0.9 or more and slope within 1±0.3; RBC count: r = 0.8 or more and slope within 1±0.3 |
| Automated differential performed for body fluids | yes (2-part differential) |
| Tests available for research use only | — |
| Differential methods used | fluorescent flow cytometry |
| Analytical measurement range: • WBC count/RBC count | WBC: 0.04–440.00 × 103/μL/RBC: 0.02–8.60 × 106/μL |
| Precision: • WBC count/RBC count | WBC: <3.0%/RBC: <1.5% |
| Accuracy: • WBC count/RBC count | — |
| Interfering substances: • WBC/RBC | no significant interference up to 39.4 mg/dL for bilirubin C, 37.4 mg/dL for bilirubin F, 996 mg/dL for hemolysis, more/no significant interference up to 39.4 mg/dL for bilirubin C, 37.4 mg/dL for bilirubin F, 996 mg/dL for hemolysis, more |
| • Hemoglobin/Platelets | no significant interference up to 39.4 mg/dL for bilirubin C, 37.4 mg/dL for bilirubin F, more/no significant interference up to 39.4 mg/dL for bilirubin C, 37.4 mg/dL for bilirubin F, 996 mg/dL for hemolysis, more |
| • Hematocrit/Reticulocytes | no significant interference up to 39.4 mg/dL for bilirubin C, 37.4 mg/dL for bilirubin F, 996 mg/dL for hemolysis, more/no significant interference up to 39.4 mg/dL for bilirubin C, 37.4 mg/dL for bilirubin F, 1, 010 mg/dL for hemolysis, more |
| • Automated differential | — |
| Throughput: max. CBCs per hour/Max. CBCs with differentials per hour | 60/60 |
| • Max. No. of slides processed per hour if stainer included | — |
| FDA-cleared or -approved sample container types (besides potassium EDTA) | none |
| Min. specimen volume open mode/Closed mode/Microsample capability | 25 µL/25 µL/yes |
| Types of patient data archived/Max. No. of archived patient data accessible when analyzer online | demographics, results, flags/10, 000 results |
| Analyzer performs delta checks | yes |
| Parameters for which flags may appear | flags for all reportable parameters deemed abnormal per lab’s protocol, more |
| Analyzer recognizes and holds test results | yes, for potential test rerun |
| Scattergram display uses cell-specific color | yes |
| Histogram display uses color with thresholds | yes |
| LIS interface formats supported | ASTM 1381, ASTM 1394-97 |
| • Middleware interface | yes (can use company's middleware or Data Innovations) |
| • Information transferred to LIS via middleware or LIS interface | numeric results, flagged results, histograms, scatterplots, patient demographics, patient orders, host query for patient demographics and orders |
| Analyzer transmits LOINC codes for results to LIS or middleware | yes |
| Lab automation system or workcells to which analyzer has a direct interface | none |
| Barcode symbologies read on specimen tube | Codabar, Code 39, Code 128, ITF, NW-7, ISBT 128, more |
| • Analyzer allows barcode placement per CLSI standard Auto02-A2 | yes |
| No. of onboard reagents required for standard specimen analysis | 4 |
| • Reagents ready to use/Reagent tracking method | yes/RFID, barcode reader |
| Onboard diagnostics for troubleshooting | yes |
| • Manufacturer can perform diagnostics via remote access | yes |
| Warranty provided/Cost of annual service contract (24 h/7 d) | one-year warranty/— |
| Notes | †also marketed by McKesson, Henry Schein |