Webinars and Sponsored Roundtables — Register Now
Tuesday, July 21, 2026, 11:00-11:30 AM CT
- Explain how transparency and manufacturer partnerships improve quality, consistency, and decision-making confidence in specimen management.
- Evaluate blood collection tubes beyond cost and commodity assumptions, incorporating clinical impact and risk into decision-making.
- Assess the potential risk points when using a blood collection device that has not been cleared for a specific purpose.
Roundtable presenters Nick Fingland, PhD, PMP, Senior Director, R&D Operations and Science, BD, and Chris Farnsworth, PhD, D(ABCC), Section Head of Clinical Chemistry, Professor of Pathology and Immunology, Washington University School of Medicine.
Moderated by: Bob McGonnagle, Publisher, CAP TODAY
Wednesday, July 29, 2026, 1:00-2:00 PM ET
Learn about digital pathology technology that is future-ready, yet practical for today’s
laboratory needs.
Webinar presenters Scott Hammond, Senior Systems Consultant, Digital Pathology Division, Wexner Medical Center, Department of Pathology, and Ursula Hofer, Imaging Technologist, Pathology Digital Imaging Lab, Wexner Medical Center, Department of Pathology, and Sandra Banky, PA(ASCP), Director of Operations, Wexner Medical Center, Department of Pathology.
Moderated by: Bob McGonnagle, Publisher, CAP TODAY
Interactive Product Guides
Product Spotlight
Beckman Coulter
DxH 520 Hematology Analyzer
| Company | Beckman Coulter |
|---|---|
| Name of analyzer | DxH 520 Hematology Analyzer |
| Contact | Eric Pabon |
| City, State | Brea, CA |
| Phone | 714-993-5321 |
| Website | 714-993-5321 http://www.beckmancoulter.com |
| Analyzer application | diagnostic and research use |
| Dimensions of analyzer (H × W × | 16 × 10.6 × 16.9 in./5.02 sq. ft. |
| Country where analyzer was | U.S./U.S. |
| First year analyzer was sold in U.S./Approximate No. of units installed in U.S.* | 2019/1, 250 |
| First year analyzer sold outside | 2019/>2, 015 |
| Foreign locations where company | worldwide (installed worldwide) |
| Human languages (other than | Japanese, Chinese, French, Italian, German, Spanish, Portuguese, Danish, Swedish, Greek, Czech, others |
| Analyzer list price | $40, 000 |
| • Analyzer acquisition options | purchase, lease |
| Training included with purchase/Total time for basic training per operator | 30 min. daily |
| • Follow-up training available | yes (extra charge) |
| Parameters that analyzer tests and reports | %&# eosinophils, %&# lymphocytes, %&# monocytes, %&# neutrophils, %&# basophils, Hb, Hct, MCH, MCHC, MCV, PLT, RBC, WBC |
| Types of testing FDA cleared or approved for analyzer | blood count analysis |
| Types of body fluids FDA cleared or approved for analysis on analyzer | — |
| Analytical measurement range for body fluid | — |
| Precision for body fluid | — |
| Accuracy for body fluid | — |
| Automated differential performed for body fluids | no |
| Tests available for research use only | low hemoglobin density, microcytic anemia factor, more |
| Differential methods used | CBC analysis based on Coulter Principle; uses simultaneous measurements of volume, axial light loss within WBC aperture to count and size lymphocytes, monocytes, neutrophils, eosinophils, basophils |
| Analytical measurement range: • WBC count/RBC count | WBC: 0.20–100.00 × 103 cells/μL/RBC: 0.20–8.00 × 106 cells/μL |
| Precision: • WBC count/RBC count | WBC: 0.20 to <1.00 × 103 cells/μL: ≤0.15 SD; 1.00 to 5.00–7.00 × 103 cells/μL: ≤4.00% CV; >7.00–100.00 × 103 cells/μL: ≤3.00% CV/RBC: 1.00–3.50 × 106 cells/μL: ≤3.00% CV; 3.50–8.00 × 106 cells/μL: ≤2.00% CV |
| Accuracy: • WBC count/RBC count | WBC: 0.20–100.00 × 103 cells/μL/RBC: 0.20–8.00 × 106 cells/μL |
| Interfering substances: • WBC/RBC | unlysed RBCs, NRBCs, cryoglobulin, cryofibrinogen, platelet clumps, giant platelets, agglutinated WBCs/agglutinated RBCs, unlysed RBCs, elevated WBCs, more |
| • Hemoglobin/Platelets | lipids >62.5 mg/dL (lipemia)/giant platelets, platelet clumps, microcytic RBCs, cryoglobulin, WBC or RBC fragments |
| • Hematocrit/Reticulocytes | agglutinated RBCs, unlysed RBCs, elevated WBCs, more/— |
| • Automated differential | unlysed RBCs, NRBCs, cryoglobulin, cryofibrinogen, platelet clumps, giant platelets, agglutinated WBCs |
| Throughput: max. CBCs per hour/Max. CBCs with differentials per hour | 60 open vial, 55 closed vial/60 open vial, 55 closed vial |
| • Max. No. of slides processed per hour if stainer included | — |
| FDA-cleared or -approved sample container types (besides potassium EDTA) | none |
| Min. specimen volume open mode/Closed mode/Microsample capability | 16.7 μL/16.7 μL/yes |
| Types of patient data archived/Max. No. of archived patient data accessible when analyzer online | results, flags, demographics/30, 000 |
| Analyzer performs delta checks | yes |
| Parameters for which flags may appear | flagging limits, system messages, editing of parameters |
| Analyzer recognizes and holds test results | yes, for follow-up, confirmatory testing, and potential test rerun (parameters for flags to hold samples defined by client or vendor) |
| Scattergram display uses cell-specific color | yes |
| Histogram display uses color with thresholds | yes |
| LIS interface formats supported | proprietary (analyzer or vendor specific), ASTM 1394-91, ASTM 1238-95, IEEE MIB |
| • Middleware interface | no (connects directly to LIS) |
| • Information transferred to LIS via middleware or LIS interface | numeric results, flagged results, histograms, scatterplots, patient demographics, patient orders |
| Analyzer transmits LOINC codes for results to LIS or middleware | yes (also provides list of machine codes for each test and corresponding LOINC code) |
| Lab automation system or workcells to which analyzer has a direct interface | none |
| Barcode symbologies read on specimen tube | Codabar, Code 39, Code 128, Interleaved 2 of 5 |
| • Analyzer allows barcode placement per CLSI standard Auto02-A2 | no |
| No. of onboard reagents required for standard specimen analysis | 3 |
| • Reagents ready to use/Reagent tracking method | yes/barcode reader |
| Onboard diagnostics for troubleshooting | no |
| • Manufacturer can perform diagnostics via remote access | no |
| Warranty provided/Cost of annual service contract (24 h/7 d) | one-year warranty/variable |
| Notes | †also marketed by Medline, McKesson, Henry Schein, Thermo Fisher Scientific |