Webinars and Sponsored Roundtables — Register Now
Thursday, August 20, 2026. 12 PM-1:30 PM ET
Hear from an expert in HER2 IHC whose work has contributed to current scoring guidelines.
Webinar presenter Josef Ruschoff, MD, PhD, Former Chief Medical Officer and Co-Founder of Targos, Senior Consultant of Discovery Life Sciences, Kassel, Germany.
Moderated by: Bob McGonnagle, Publisher, CAP TODAY
To be rescheduled to a future date
Learn about digital pathology technology that is future-ready, yet practical for today’s
laboratory needs.
Webinar presenters Scott Hammond, Senior Systems Consultant, Digital Pathology Division, Wexner Medical Center, Department of Pathology, and Ursula Hofer, Imaging Technologist, Pathology Digital Imaging Lab, Wexner Medical Center, Department of Pathology, and Sandra Banky, PA(ASCP), Director of Operations, Wexner Medical Center, Department of Pathology.
Moderated by: Bob McGonnagle, Publisher, CAP TODAY
Interactive Product Guides
Product Spotlight
Sysmex Corp.
XQ-320
| Company | Sysmex Corp. |
|---|---|
| Name of analyzer | XQ-320 |
| Contact | Madelaine Dintelman |
| City, State | Lincolnshire, IL |
| Phone | 888-879-7639 |
| Website | |
| Analyzer application | diagnostic use |
| Dimensions of analyzer (H × W × | 17 × 14 × 17.5 in./— |
| Country where analyzer was | Japan/Japan |
| First year analyzer was sold in U.S./Approximate No. of units installed in U.S.* | 2025/— |
| First year analyzer sold outside | — |
| Foreign locations where company | — |
| Human languages (other than | — |
| Analyzer list price | $33, 998.80 (slide maker stainer sold separately) |
| • Analyzer acquisition options | purchase, trade in, lease |
| Training included with purchase/Total time for basic training per operator | 2 min. daily; 15 min. weekly |
| • Follow-up training available | yes (no extra charge) |
| Parameters that analyzer tests and reports | %&# lymphocytes, %&# neutrophils, Hb, Hct, MCH, MCHC, MCV, PLT, RBC, WBC |
| Types of testing FDA cleared or approved for analyzer | — |
| Types of body fluids FDA cleared or approved for analysis on analyzer | — |
| Analytical measurement range for body fluid | — |
| Precision for body fluid | — |
| Accuracy for body fluid | — |
| Automated differential performed for body fluids | — |
| Tests available for research use only | — |
| Differential methods used | direct current |
| Analytical measurement range: • WBC count/RBC count | WBC: 0.20–99.9 × 103/μL/RBC: 0.01–7.0 × 106/μL |
| Precision: • WBC count/RBC count | WBC: <3.5%/RBC: <2.0% |
| Accuracy: • WBC count/RBC count | WBC: within ±3% or ±0.20 × 103/μL/RBC: within ±2% or ±0.03 × 106/μL |
| Interfering substances: • WBC/RBC | cold agglutinins, platelet clumps, cryoprotein, cryoglobulin, fibrin, giant platelets (>1 M/µL)/cold agglutinins, severe microcytosis, fragmented RBCs, leukocytosis (>100, 000/µL), giant platelets (>1 M/µL) |
| • Hemoglobin/Platelets | severe lipemia, abnormal protein, leukocytosis (>100, 000/µL)/platelet clumps, pseudothrombocytopenia, giant platelets, severe microcytosis, fragmented RBCs, fragmented leukocytes, cryoprotein, cryoglobulin |
| • Hematocrit/Reticulocytes | cold agglutinins, severe microcytosis, fragmented RBCs, leukocytosis (>100, 000/µL), severe diabetes, uremia, spherocytosis/— |
| • Automated differential | — |
| Throughput: max. CBCs per hour/Max. CBCs with differentials per hour | 70/70 |
| • Max. No. of slides processed per hour if stainer included | — |
| FDA-cleared or -approved sample container types (besides potassium EDTA) | none |
| Min. specimen volume open mode/Closed mode/Microsample capability | 16 µL/—/yes |
| Types of patient data archived/Max. No. of archived patient data accessible when analyzer online | demographics, results, flags/10, 000 |
| Analyzer performs delta checks | no |
| Parameters for which flags may appear | WBC histogram, RBC histogram, platelet histogram, error flags |
| Analyzer recognizes and holds test results | yes, for potential test rerun (parameters for flags to hold samples defined by vendor) |
| Scattergram display uses cell-specific color | no |
| Histogram display uses color with thresholds | yes |
| LIS interface formats supported | ASTM 1381, HL7 version 2, ASTM E1394-97 |
| • Middleware interface | yes (can use company’s middleware or third-party middleware) |
| • Information transferred to LIS via middleware or LIS interface | numeric results, flagged results, histograms, scatterplots, patient demographics, patient orders, host query for patient demographics and orders |
| Analyzer transmits LOINC codes for results to LIS or middleware | yes |
| Lab automation system or workcells to which analyzer has a direct interface | none |
| Barcode symbologies read on specimen tube | Codabar, Code 39, Code 128, ITF, NW7, ISBT 128, JAN/EAN/UPC |
| • Analyzer allows barcode placement per CLSI standard Auto02-A2 | yes (supports QR codes) |
| No. of onboard reagents required for standard specimen analysis | 2 |
| • Reagents ready to use/Reagent tracking method | yes/RFID, barcode reader |
| Onboard diagnostics for troubleshooting | yes |
| • Manufacturer can perform diagnostics via remote access | yes |
| Warranty provided/Cost of annual service contract (24 h/7 d) | one-year warranty/— |
| Notes | †also marketed by McKesson, Henry Schein |