Webinars and Commercial Conversations — Register Now

Tuesday, November 10, 2026, 1:00 PM-2 PM ET
Join us for a session to help improve diagnostic accuracy in your practice when scoring HER2 IHC 0 with faint, incomplete membrane staining in ≤10% of tumor cells vs IHC 0 with no membrane staining detected in patients with mBC.

Webinar presenter Samuel Caughron, MD, FCAP, Chief Strategy Officer, MAWD Pathology, Medical Director of Laboratory, AdventHealth Kansas City, Medical Director of Laboratory, Labcorp, President, Member & Director, Pathology Ventures, LLC

CAP TODAY does not endorse any of the products or services named within. This program is intended for US healthcare professionals. This program is being sponsored by AstraZeneca. The speaker is being compensated for the presentation. The program is not CME accredited and may not be used for CME accreditation.

Moderated by: Bob McGonnagle, Publisher, CAP TODAY

Monday, November 16, 2026, 12:30 PM-2 PM ET
Hear experts discuss a multidisciplinary approach to oncopathology for HER2
assessments for solid tumors in the absence of a companion diagnostic.

Webinar presenters Funda Meric-Bernstam, MD, Chair, Department of Investigational Cancer Therapeutics Medical Director, Institute for Personalized Cancer Therapy, The University of Texas MD Anderson Cancer Center, Houston, Texas, and Emina E. Torlakovic, MD, PhD, College of Medicine, University of Saskatchewan, Canadian Biomarker Quality Assurance, Saskatoon, Saskatchewan, Canada, and Carol Cheung, MD, PhD, JD, FRCPC, Deputy Director, Canadian Biomarker Quality, Assurance—Programme canadien d’assurance de la qualité des biomarqueurs

CAP TODAY does not endorse any of the products or services named within. This program is being sponsored by Daiichi Sankyo, Inc. and AstraZeneca. The speaker is being compensated for the presentation. The program is not CME accredited and may not be used for CME accreditation.

Moderated by: Bob McGonnagle, Publisher, CAP TODAY

Wednesday, November 18, 2026, 12:00 PM-1 PM ET
Join leading pathology experts, Savitri Krishnamurthy, MD, and Marilyn Bui, MD, PhD, for a webinar exploring potential challenges in HER2 IHC testing for breast cancer, emerging evidence supporting digital and computational pathology as assistive tools for pathologists in HER2 IHC assessment, and practical considerations for real-world implementation and adoption.

Webinar presenters Savitri Krishnamurthy, MD, Professor of Pathology, Department of Anatomical Pathology The University of Texas MD Anderson Cancer Center, Houston, Texas, and Marilyn Bui, MD, PhD, Surgical Pathologist and Cytopathologist Member and Professor, Departments of Pathology, Machine Learning, and Translational Pathology, Scientific Director, Analytic Microscopy Core Chair, Women in Oncology Executive Board Moffitt Cancer Center, Tampa, Florida

CAP TODAY does not endorse any of the products or services named within. This program is being sponsored by Daiichi Sankyo, Inc. and AstraZeneca. The speakers are being compensated for their presentations. The program is not CME accredited and may not be used for CME credit.

Moderated by: Bob McGonnagle, Publisher, CAP TODAY

Wednesday, December 9, 2026, 12 PM-1 PM ET
Hear an expert pathologist discuss the role of HER2 IHC testing and scoring in solid tumors including endometrial, ovarian, cervical, bladder, non-small cell lung, colorectal, and biliary cancers

Webinar presenter Michelle Shiller, DO, MSPT, Baylor Sammons Cancer Center; Fulgent Genetics; Texas Cancer Interception Institute

CAP TODAY does not endorse any of the products or services named within. This program is intended for US healthcare professionals. This program is being sponsored by AstraZeneca. The speaker is being compensated for the presentation. The program is not CME accredited and may not be used for CME accreditation.

Moderated by: Bob McGonnagle, Publisher, CAP TODAY

Subspecialties

Interactive Product Guides

Manage Subscription

To receive CAP TODAY magazine, please subscribe here.
For change of address, mail new information to: CAP TODAY Circulation, 325 Waukegan Road, Northfield, IL 60093-2750 or visit here.

Product Spotlight

Sysmex Corp.

XR-Series

Company

Sysmex Corp.

Name of analyzer

XR-Series

Contact

Jill Crist  

Email

communications@sysmex.com

City, State

Lincolnshire, IL

Phone

888-879-7639  

Website

http://www.sysmex.com

Analyzer application

diagnostic use

Dimensions of analyzer (H × W × D)/Analyzer footprint 

33.7 × 25.4 × 29.7 in./5.24

Country where analyzer was designed/Manufactured 

Japan/Japan

First year analyzer was sold in U.S./Approximate No. of units installed in U.S. (based on August 2026 survey deadline) 

2026/0

First year analyzer sold outside U.S./Approximate No. of units installed outside U.S. (based on August 2026 survey deadline) 

2024/4,360

Foreign locations where company markets analyzer

all countries except Latin America region (installed in all countries except Latin America region)

Human languages (other than English) supported by analyzer 

—

Analyzer list price

$165,000 (XR-1000-100; slide maker stainer sold separately)

• Analyzer acquisition options

purchase, trade in, lease

Training included with purchase/Total time for basic training per operator

yes/—(at customer site)

• Follow-up training available

yes (no extra charge)

Parameters that analyzer tests and reports

%&# eosinophils, %&# immature granulocytes, %&# lymphocytes, %&# monocytes, %&# neutrophils, %&# basophils, %&# reticulocyte count, Hb, Hct, MCH, MCHC, MCV, PLT, RBC, nucleated red blood cells, WBC

Types of testing FDA cleared or approved for analyzer

blood count analysis, body fluid analysis

Types of body fluids FDA cleared or approved for analysis on analyzer

CSF, peritoneal/ascites, pleural fluid, synovial fluid

Analytical measurement range for body fluid

WBC count: 0.003–10.000 × 103/μL; RBC count: 0.002–5.000 × 106/μL; TNC count: 0.003–10.000 × 103/μL

Precision for body fluid

WBC count: ≤30.0% (0.005–0.015 × 103/μL), ≤15.0% (0.016–0.030 × 103/μL), ≤10.0% (0.031–0.050 × 103/μL); RBC count: ≤40.0% or ≤0.007 × 106/μL (0.003–0.050 × 106/μL) max.–min.; TNC count: ≤30.0% (0.005–0.015 × 103/μL), ≤15.0% (0.016–0.030 × 103/μL), ≤10.0% (0.031–0.050 × 103/μL)

Accuracy for body fluid

—

Automated differential performed for body fluids

yes (2-part differential)

Tests available for research use only

microR, macroR, PLT-O, RBC-O, RBC-He, PDW, PCT, more

Differential methods used

fluorescent flow cytometry

Analytical measurement range: • WBC count/RBC count

WBC: 0.03–440.00 × 103/μL/RBC: 0.01–8.60 × 106/μL

Precision: • WBC count/RBC count

WBC: ≤3.0%/RBC: ≤1.5%

Accuracy: • WBC count/RBC count

WBC: within ±3% or ±0.20 × 103/μL/RBC: within ±2% or ±0.03 ×106/μL 

Interfering substances: • WBC/RBC

WBC aggregation, possibility of platelet clumps, cryoprotein, cryoglobulin, fibrin, giant platelets/ erythrocyte aggregation (cold agglutinin), microerythrocytes, possibility of fragmented RBCs

• Hemoglobin/Platelets

leukocytosis, lipemia, abnormal protein/possibility of platelet clumps, pseudothrombocytopenia, giant platelets, microerythrocytes, possibility of fragmented RBCs, more

• Hematocrit/Reticulocytes

erythrocyte aggregation (cold agglutinin), microerythrocytes, possibility of fragmented RBCs, leukocytosis, more/erythrocyte aggregation (cold agglutinin), giant platelets, possibility of platelet clumps, fragmented leukocytes, more

• Automated differential

—

Throughput: max. CBCs per hour/Max. CBCs with differentials per hour

110/110

• Max. No. of slides processed per hour if stainer included

70

FDA-cleared or -approved sample container types (besides potassium EDTA)

—

Min. specimen volume open mode/Closed mode/Microsample capability

300 μL/1 mL/yes

Types of patient data archived/Max. No. of archived patient data accessible when analyzer online

patient demographics, test results, flags, scattergrams, histograms/100,000

Analyzer performs delta checks

yes

Parameters for which flags may appear

flags for all reportable parameters deemed abnormal per lab’s protocol, more

• Flagging client selectable

combination of client and vendor selectable

Analyzer recognizes and holds test results

yes, for confirmatory and potential test rerun (parameters for flags to hold samples defined by client or vendor)

Scattergram display uses cell-specific color

yes

Histogram display uses color with thresholds

no

LIS interface formats supported

ASTM 1394-91, ASTM 1238-95,  ASTM 1381

• Middleware interface

yes (can use company’s middleware or Data Innovations)

Analyzer transmits LOINC codes for results to LIS or middleware

yes (also provides list of machine codes for each test and corresponding LOINC code)

Lab automation system or workcells to which analyzer has a direct interface

Beckman Coulter, Siemens Healthineers, QuidelOrtho, Sysmex, Thermo Fisher, Roche, Abbott

Barcode symbologies read on specimen tube

Codabar (NW-7), Code 39, Code 128, JAN/EAN/UPC

• Analyzer allows barcode placement per CLSI standard Auto02-A2

yes

No. of onboard reagents required for standard specimen analysis

8

• Reagents ready to use/Reagent tracking method

yes/RFID, barcode reader

Time required for analyzer maintenance

20 min. daily; 20 min. weekly

Onboard diagnostics for troubleshooting

yes

• Manufacturer can perform diagnostics via remote access

yes

Warranty provided/Cost of annual service contract (24 h/7 d)

one-year warranty/$31,807.92 (XR-1000-100)

Distinguishing analyzer features (supplied by company)

high-throughput analyzer with enhanced cell population differentiation; 3D scattergrams give deeper insight into abnormal samples

Note

†also marketed by McKesson, Henry Schein

More products
in this guide

Alinity h-series
CELL-DYN Emerald
CELL-DYN Emerald 22
CELL-DYN Emerald 22 Autoloader
DxH 500 Hematology Analyzer
DxH 520 Hematology Analyzer
DxH 560 AL Hematology Analyzer
DxH 690T Hematology Analyzer
DxH 900 Workcell Automated Hematology Solution
ABX Micros 60
ABX Micros ES60
ABX Pentra 60C+
ABX Pentra XL80
Yumizen H2500
GloCyte Automated Cell Counter for CSF
X100/X100HT with Full-Field Peripheral Blood Smear Application
XN-330, XN-430, XN-530
XN-350, XN-450, XN-550
XN Series (XN-1000, XN-2000, XN-3100, XN-9100)