Webinars and Sponsored Roundtables — Register Now

Tuesday, July 21, 2026, 11:00-11:30 AM CT

Learning Objectives:
  • Explain how transparency and manufacturer partnerships improve quality, consistency, and decision-making confidence in specimen management.
  • Evaluate blood collection tubes beyond cost and commodity assumptions, incorporating clinical impact and risk into decision-making.
  • Assess the potential risk points when using a blood collection device that has not been cleared for a specific purpose.

Roundtable presenters Nick Fingland, PhD, PMP, Senior Director, R&D Operations and Science, BD, and Chris Farnsworth, PhD, D(ABCC), Section Head of Clinical Chemistry, Professor of Pathology and Immunology, Washington University School of Medicine.

Moderated by: Bob McGonnagle, Publisher, CAP TODAY

Wednesday, July 29, 2026, 1:00-2:00 PM ET
Learn about digital pathology technology that is future-ready, yet practical for today’s
laboratory needs.

Webinar presenters Scott Hammond, Senior Systems Consultant, Digital Pathology Division, Wexner Medical Center, Department of Pathology, and Ursula Hofer, Imaging Technologist, Pathology Digital Imaging Lab, Wexner Medical Center, Department of Pathology, and Sandra Banky, PA(ASCP), Director of Operations, Wexner Medical Center, Department of Pathology.

Moderated by: Bob McGonnagle, Publisher, CAP TODAY

Subspecialties

Interactive Product Guides

Product Spotlight

Beckman Coulter

Unicel DxI 600

Company

Beckman Coulter

Name of instrument

Unicel DxI 600

Contact

Doug Lee   

Email

dylee@beckman.com

City, State

Brea, CA

Phone

800-526-3821   

Website

http://www.beckmancoulter.com

Type of instrument

immunoassay

List price/First year sold in U.S.

—/2006

Targeted hospital bed size/Targeted test volume

Company manufactures instrument

yes (also sold by McKesson, Henry Schein, Medline, Thermo Fisher Scientific)

Other models in this family of analyzers

Access 2, Unicel DxI 800

No. of units in clinical use in U.S./Outside U.S. (countries)

Dimensions (H × W × D)/Instrument footprint

67 × 61.5 × 37.5 in./16 sq. ft.

Weight empty/Weight fully loaded

1,065 lbs./—

No. of different measured assays onboard simultaneously

50 (50 can be run and calibrated at one time)

No. of user-definable (open chemistry) channels

0

Test throughput per hour/Assay run time

up to 200/13–55 min.

No. of direct ion-selective electrode channels

Detection methods

Stat time until completion/specimen throughput for Ion-selective electrode

Stat time until completion/specimen throughput for Ion-selective electrode basic metabolic panel

Stat time until completion/specimen throughput for Ion-selective electrode complete metabolic panel

Typical time delay from ordering stat test until aspiration of sample

Fully automated microplate immunoassay system

no

Methodologies supported

chemiluminescence

Separation methodologies

magnetic particle

Stat time until completion of a ß-hCG test

15 min.

Stat time until completion of a cTn test

17 min.

• Typical time delay from test order to aspiration of sample

18 sec.

Reagents refrigerated onboard/Reagents ready to use

yes (4°–10°C)/yes

Reagent lot tracking/Reagent inventory

yes/yes

Reagent form/Reagents barcoded

liquid chemistry (closed reagent system)/yes

Separate reagent pack for each specimen/for each test run

no/no

Walkaway capability/Walkaway duration

yes/180 min. or 60 specimens

Design of sample-handling system

rack

Uses washable cuvettes/Uses disposable cuvettes

no/yes (can store > 1,000 cuvettes)

Min.–max. sample volume that can be aspirated at one time 

5–200 µL

Min. reaction volume/Min. specimen volume/Min. dead volume

10 µL/150 µL/140 µL

Dedicated pediatric sample cup

yes (dead volume: 100 µL)

Primary tube sampling

yes

Accommodates most standard tube sizes/Accepts nonstandard tube sizes

yes/—

Pierces caps on primary tubes

no

Protects against probe collision

no

Detects clots/liquid level/short sample

yes/yes/yes

Detection or quantitation for hemolysis, icterus, lipemia, clots

detection for clots; hemolysis, icterus, lipemia not available

Dilutes patient samples onboard/Susceptibility to carryover

yes (can be programmed to perform dilutions prior to analysis)/—

Automatic rerun capability

yes

Sample volume can be diluted to rerun out-of-linear-range high results

yes

Sample volume can be concentrated to rerun out-of-linear-range low results

no

Autocalibration/Multipoint calibration supported

no (calibrants are not stored onboard)/yes (recommended avg. frequency: 28 days)

Typical calibration frequency for ISE/therapeutic drugs/drugs of abuse/general chemistries/immunoassays

—/—/—/—/28 days

Automatic programmable start/Automatic programmable shutdown

no/no

Onboard real-time QC/Onboard software capability to review QC

no/yes

Supports multiple QC lot numbers per analyte

yes

Waste management

direct to drain

Sample barcode-reading capability/Autodiscrimination

yes (Interleaved 2 of 5, Codabar, Code 39, Code 128)/no

Lab can control analyzer from remote computer

yes

Instrument can diagnose its own malfunctions

yes (operator intervention required to order parts)

System malfunctions can be diagnosed via remote monitoring

yes

UPS backup power supply

yes

Data-management capability/LIS or EHR systems interfaced

onboard/Oracle Health, Antrim, CCA, Chemware, Dawning Technologies, SCC, Dynamic Healthcare, Antek, more

LIS interface provided/Bidirectional interface capability

yes (included in instrument price)/yes (broadcast download and host query)

Modem servicing provided/Service engineer on-site response time

yes/within 24 hrs.

Mean time between failures

3.1 down service calls per year (displays error codes for troubleshooting)

Average scheduled maintenance time by lab personnel

daily: < 10 min.; weekly: 15 min. at 5, 000 tests; monthly: 35 min. at 10, 000 tests

Maintenance records kept onboard for user/vendor

yes/no

Maintenance training demonstration module onboard

yes

Training included with purchase/Avg. time for basic user training

yes (2 training slots)/3 days (at vendor site)

Advanced operator training/Extra charge for follow-up or advanced training

yes (at vendor site)/yes

Warranty provided/Cost of annual service contract (24 h/7 d)

yes (1 year)/—

Distinguishing features (supplied by company)

onboard aliquoting quickly frees samples for other analyses; scalable results across all immunoassay systems; liquid, ready-to-use reagents

More products
in this guide

Alinity ci-series
ASI Evolution RPR Syphilis Analyzer
DxI 9000 Access Immunoassay Analyzer
Pictus 700 (P700)
Agility Automated ELISA System
AIX1000 Agglutination Instrument
Yumizen C1200
Yumizen C560
VITROS 3600 Immunodiagnostic System
VITROS 4600 Chemistry System
VITROS XT 7600 Integrated System
cobas pro integrated solutions
dIFine P30 Processor
IMMULITE 2000 XPi Immunoassay System
Phadia 250 Laboratory System
B⋅R⋅A⋅H⋅M⋅S KRYPTOR GOLD
QUANTA-Lyser 3000